Details
Measurement of whole-plant gas exchange (photosynthesis, transpiration and stomatal conductance) in controlled environmental conditions for small plants like A.thaliana.
System allows for rapid modification and application of:
CO2 concentration - addition or removal
Relative air humidity (VPD)
Ozone
Light intensity and spectra
Exogenous spray treatment of various substances (hormones, chemicals, nanoparticles etc.)
Stimuli ranges (systemic)
Relative air humidity: 30-80%
Air temperature 5-40° C
Stimuli ranges (per chamber)
CO2 concentration: 5-1000 ppm
Ozone concentration: 10-1500 ppb
White light intensity: 0-1000 µmol m-2s-1
Red light intensity: 0-1200 µmol m-2s-1
Blue light intensity: 0-1800 µmol m-2s-1
Green light intensity: 0-1000 µmol m-2s-1
Cuvette dimensions
Width: 77 mm
Height: 36 mm
Volume: 167.64 cm3
Data sampling rate
1 min per cuvette: 30 sec reference + 30 sec sample
Air flow rate: 20ml/s
Measurement accuracy
CO2 concentration: +- 0.1 ppm
H2O concentration: +- 0.1 ppm
Air temperature: +- 0.1 C
Ozone concentration: +- 10 ppb
Flow meter: 0.5 ml/s
System dimensions
2×1 metres
System requires access to building ventilation