Details

Measurement of whole-plant gas exchange (photosynthesis, transpiration and stomatal conductance) in controlled environmental conditions for small plants like A.thaliana.

System allows for rapid modification and application of:

  • CO2 concentration - addition or removal

  • Relative air humidity (VPD)

  • Ozone

  • Light intensity and spectra

  • Exogenous spray treatment of various substances (hormones, chemicals, nanoparticles etc.)

Stimuli ranges (systemic)

  • Relative air humidity: 30-80%

  • Air temperature 5-40° C

Stimuli ranges (per chamber)

  • CO2 concentration: 5-1000 ppm

  • Ozone concentration: 10-1500 ppb

  • White light intensity: 0-1000 µmol m-2s-1

  • Red light intensity: 0-1200 µmol m-2s-1

  • Blue light intensity: 0-1800 µmol m-2s-1

  • Green light intensity: 0-1000 µmol m-2s-1

Cuvette dimensions

  • Width: 77 mm

  • Height: 36 mm

  • Volume: 167.64 cm3

Data sampling rate

  • 1 min per cuvette: 30 sec reference + 30 sec sample

  • Air flow rate: 20ml/s

Measurement accuracy

  • CO2 concentration: +- 0.1 ppm

  • H2O concentration: +- 0.1 ppm

  • Air temperature: +- 0.1 C

  • Ozone concentration: +- 10 ppb

  • Flow meter: 0.5 ml/s

System dimensions

  • 2×1 metres

System requires access to building ventilation